Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 31: CAL ADJ /5520 INSTRUMENT: Fluke 33: CAL ADJ /5520 DATE: 18-Feb-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 59 CONFIGURATION: Fluke 5520A STANDARD: Fluke 5500A/COIL ============================================================================= # # Source: # Fluke 31/33 Instruction Manual (PN 896118 Sep. 1991). # Fluke 5520A/COIL Instruction Sheet (PN 600616 Aug. 1995) # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5520A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 STD Fluke 5500A/COIL 1.002 ASK- W 1.003 HEAD EQUIPMENT SETUP 1.004 DISP This procedure requires a 50-turn Toroid Coil. 1.004 DISP Refer to the 31/33 Instruction Manual figure 7, page 32 1.004 DISP for directions for constructioning a suitable coil. 1.005 HEAD CALIBRATION ADJUSTMENT 1.006 DISP Use a Phillips-head screwdriver to remove the back from 1.006 DISP the 30-Series meter to be calibrated. Wrap a rubberband 1.006 DISP around the case and battery to keep the battery in 1.006 DISP place. Note the location of the calibration point. 1.007 DISP Connect the 5500A/COIL to the 5520A 20A & AUX Lo 1.007 DISP terminals. 1.007 DISP 1.007 DISP Clamp the meter around the toroid center conductor 1.007 DISP bundle, positioning at the alignment marks. 1.008 DISP Insert a small insulated screwdriver into the 1.008 DISP calibration potentiometer and turn the UUT on. 1.009 5520 300.0A 50H SI HC S LT5 1.010 DISP Adjust the calibration potentiometer for a measurement 1.010 DISP of 300.0A. 1.011 END